Blood pressure measurement Presented by -Deepika Jain -Divya Jain

Blood pressure Normal blood pressure Blood pressure of 140/90 Blood pressure (BP) is a force exerted by circulating blood on the walls of blood vessels. During each heartbeat, BP varies between a maximum (systolic) and a minimum (diastolic) pressure. Normal blood pressure Blood pressure of 140/90

METHODS FOR BLOOD PRESSURE MEASUREMENT DIRECT METHOD INDIRECT METHOD

INDIRECT METHOD OF BLOOD PRESSURE MEASURMENT KOROTOKOFF SOUNDS METHOD AUSCULTATORY METHOD OSICILLOMETRIC METHOD RHEOGRAPHIC METHOD

KOROTOKOFF SOUNDS Korotkoff are the sounds that medical personnel listen for when they are taking blood pressure using a non-invasive procedure. If the pressure is dropped to a level equal to that of the patient's systolic blood pressure, the first Korotkoff sound will be heard. As the pressure in the cuff is the same as the pressure produced by the heart, some blood will be able to pass through the upper arm when the pressure in the artery rises during systole.

The sounds heard during measurement of blood pressure are not the same as the heart sounds 'lub' and 'dub' that are due to the closing of the hearts valves. If a stethoscope is placed over the brachial artery in the antecubital fossa in a normal person (without arterial disease), no sound should be audible. As the heart beats, these pulses are transmitted smoothly via laminar (non-turbulent) blood flow throughout the arteries and no sound is produced. Also, if the cuff of a sphygmomanometer is placed around a patient's upper arm and inflated to a pressure above the patient's systolic blood pressure, there will be no sound audible. This is because the pressure in the cuff is high enough such that it completely occludes the blood flow. It is similar to a flexible tube or pipe with fluid in it that is being pinched shut. .

As the pressure in the cuff is allowed to fall further, thumping sounds continue to be heard as long as the pressure in the cuff is between the systolic and diastolic pressures, as the arterial pressure keeps on rising above and dropping back below the pressure in the cuff. Eventually, as the pressure in the cuff drops further, the sounds change in quality, then become muted, then disappear altogether. As the pressure in the cuff drops below the diastolic blood pressure, the cuff no longer provides any restriction to blood flow allowing the blood flow to become smooth again with no turbulence and thus produce no further audible sound

THE FIVE KOROTOKOFF SOUNDS Korotkoff actually described five types of Korotkoff sounds: The first Korotkoff sound is the snapping sound first heard at the systolic pressure. The second sounds are the murmurs heard for most of the area between the systolic and diastolic pressures. The third = A loud, crisp tapping sound. The fourth sound, at pressures within 10 mmHg above the diastolic blood pressure, were described as "thumping" and "muting". The fifth Korotkoff sound is silence as the cuff pressure drops below the diastolic blood pressure. The second and third Korotkoff sounds haven't had clinical significance.

AUSCULATORY METHOD The auscultatory method (from the Latin word for listening) uses a stethoscope and a sphygmomanometer. This comprises an inflatable cuff placed around the upper arm at roughly the same vertical height as the heart, attached to mercury or aneroid manometer . The mercury manometer measures the height of a column of mercury, giving an absolute result without need for calibration, and consequently not subject to the errors and drift of calibration which affect other methods. The use of mercury manometers is often required in clinical trials and for the clinical measurement of hypertension in high risk patients, such as pregnant women.

A cuff of appropriate size is fitted smoothly ,then inflated manually by repeatedly squeezing a rubber bulb until the artery is completely occluded. Listening with the stethoscope to the brachial artery at the elbow, the examiner slowly releases the pressure in the cuff. When blood just starts to flow in the artery, the turbulent flow creates a "whooshing" or pounding (first Korotkoff sound). The pressure at which this sound is first heard is the systolic BP. The cuff pressure is further released until no sound can be heard (fifth Korotkoff sound), at the diastolic arterial pressure. The auscultatory method has been predominant since the beginning of BP measurements but in other cases it's being replaced by other noninvasive techniques.

Auscultatory method aneroid sphygmomanometer with stethoscope

MERCURY MANOMETER

OSCILLOMETRIC METHOD

OSCILLOMETRIC METHOD Automated method of non invasive BP measurement It has some distinct advantages over the auscultatory method Sound is not used during measurement This technique does not require a microphone or transducer in the cuff.

Disadvantage of oscillometric method as well as auscultatory method is that the excessive movement or vibration can cause inaccurate readings PRINCIPLE “occluding cuff deflates from a level above the systolic pressure,the artery walls begin to vibrate as the blood flows through the partially occluded artery and these vibrations will be sensed in the transducer system monitoring cuff pressure”

The cuff pressure at the point of maximum oscillations usually corresponds to the mean arterial pressure. The point above the mean pressure at which the oscillations begin to rapidly decease in amplitude correlates with the diastolic pressure.

RHEOGRAPHIC METHOD A fully automatic apparatus for measuring systolic and diastolic BP has been developed using the ordinary cuff and the principle of rheographic detection of an arterial pulse. The change in impedance at two points under the occluding cuff form the basis of detection of the diastolic pressure.

In this method,a set of 3 electrodes are placed in contact with the skin A good contact is essential to reduce the skin elctrode contact impedance. Electrode B which acts as a common electrode is positioned slightly distal from the midline of the cuff.

Electrode A and C are placed at a certain distance from the electrode B,one distally and other proximally. High freq current source of 100KHz at the input. When we measure the impedance b/w any two electrodes b4 pressurizing the cuff,it shows modulation in accordance with the blood flow pulsations in the artery.

ADVANTAGES In this method of measuring BP,the cuff need not be precisely positioned as in the case with the korotkoff microphone which is to be fixed exactly above an artery. Also, the readings are not affected by ambient sound.

Thank you

1. Update on the 1987 Task Force Report on High Blood Pressure in Children and Adolescents a working group report from the National High Blood Pressure Education Program. National High Blood Pressure Education Program Working Group on Hypertension Control in Children and Adolescents. Pediatrics. 1996;98:649. [PubMed] [Google Scholar]

2. Association for the Advancement of Medical Instrumentation. Electronic of automated sphygmomanometer. Arlington, VA: American National Standard; 1992. [Google Scholar]

3. Ayman P, Goldshine AD. Blood pressure determinations by patients with essential hypertension 1: the difference between clinic and home readings before treatment. American Journal of Medical Sciences. 1940;465 [Google Scholar]

4. Bailey RH, Knaus VL, Bauer JH. Aneroid sphygmomanometers. An assessment of accuracy at a university hospital and clinics. Arch Intern Med. 1991;151:1409. [PubMed] [Google Scholar]

5. Bobrie G, Chatellier G, Genes N, et al. Cardiovascular prognosis of “masked hypertension” detected by blood pressure self-measurement in elderly treated hypertensive patients. JAMA. 2004;291:1342. [PubMed] [Google Scholar]

6. Borow KM, Newburger JW. Noninvasive estimation of central aortic pressure using the oscillometric method for analyzing systemic artery pulsatile blood flow: comparative study of indirect systolic, diastolic, and mean brachial artery pressure with simultaneous direct ascending aortic pressure measurements. Am Heart J. 1982;103:879. [PubMed] [Google Scholar]

7. Burke MJ, Towers HM, O’Malley K, et al. Sphygmomanometers in hospital and family practice: problems and recommendations. Br Med J (Clin Res Ed) 1982;285:469. [PMC free article] [PubMed] [Google Scholar]

8. Canzanello VJ, Jensen PL, Schwartz GL. Are aneroid sphygmomanometers accurate in hospital and clinic settings? Arch Intern Med. 2001;161:729. [PubMed] [Google Scholar]

9. Cavallini MC, Roman MJ, Blank SG, et al. Association of the auscultatory gap with vascular disease in hypertensive patients. Ann Intern Med. 1996;124:877. [PubMed] [Google Scholar]

10. Chobanian AV, Bakris GL, Black HR, et al. The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure: the JNC 7 report. Jama. 2003;289:2560. [PubMed] [Google Scholar]

11. Clark LA, Denby L, Pregibon D, et al. A quantitative analysis of the effects of activity and time of day on the diurnal variations of blood pressure. Journal of Chronic Diseases. 1987;40:671. [PubMed] [Google Scholar]

12. Comstock GW. An epidemiologic study of blood pressure levels in a biracial community in the Southern United States. Am J Hyg. 1957;65:271. [PubMed] [Google Scholar]

13. Cooper R, Puras A, Tracy J, et al. Evaluation of an electronic blood pressure device for epidemiological studies. Blood Press Monit. 1997;2:35. [PubMed] [Google Scholar]

14. Cushman WC, Cooper KM, Horne RA, et al. Effect of back support and stethoscope head on seated blood pressure determinations. Am J Hypertens. 1990;3:240. [PubMed] [Google Scholar]

15. de Swiet M, Dillon MJ, Littler W, et al. Measurement of blood pressure in children. Recommendations of a working party of the British Hypertension Society. BMJ. 1989;299:497. [PMC free article] [PubMed] [Google Scholar]

16. Eckert S, Gleichmann S, Gleichmann U. Blood pressure self-measurement in upper arm and in wrist for treatment control of arterial hypertension compared to ABPM. Z Kardiol. 1996;85 (Suppl 3):109. [PubMed] [Google Scholar]

17. Eilertsen E, Humerfelt S. The observer variation in the measurement of arterial blood pressure. Acta Med Scand. 1968;184:145. [PubMed] [Google Scholar]

18. Elseed AM, Shinebourne EA, Joseph MC. Assessment of techniques for measurement of blood pressure in infants and children. Arch Dis Child. 1973;48:932. [PMC free article] [PubMed] [Google Scholar]

19. Fagard RH, Staessen JA, Thijs L, et al. Response to antihypertensive therapy in older patients with sustained and nonsustained systolic hypertension. Systolic Hypertension in Europe (Syst-Eur) Trial Investigators. Circulation. 2000;102:1139. [PubMed] [Google Scholar]

20. Fagard RH, Van Den Broeke C, De Cort P. Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice. J Hum Hypertens. 2005;19:801. [PubMed] [Google Scholar]

21. Finnegan TP, Spence JD, Wong DG, et al. Blood Pressure Measurement in the Elderly: Correlation of Arterial Stiffness with Difference Between Intra-arterial and Cuff Pressures. Journal of Hypertension. 1985;3:231. [PubMed] [Google Scholar]

22. Foster C, McKinlay S, Cruickshank JM, et al. Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Hum Hypertens. 1994;8:661. [PubMed] [Google Scholar]

23. Gerin W, Schwartz AR, Schwartz JE, et al. Limitations of current validation protocols for home blood pressure monitors for individual patients. Blood Press Monit. 2002;7:313. [PubMed] [Google Scholar]

24. Gould BA, Hornung RS, Altman DG, et al. Indirect measurement of blood pressure during exercise testing can be misleading. British Heart Journal. 1985;53:611. [PMC free article] [PubMed] [Google Scholar]

25. Henschel A, De La Vega F, Taylor HL. Simultaneous direct and indirect blood pressure measurements in man at rest and work. J Appl Physiol. 1954;6:506. [PubMed] [Google Scholar]

26. Imai Y, Abe K, Sasaki S, et al. Clinical evaluation of semiautomatic and automatic devices for home blood pressure measurement: comparison between cuff-oscillometric and microphone methods. Journal of Hypertension. 1989;7:983. [PubMed] [Google Scholar]

27. Imholz BP, Langewouters GJ, van Montfrans GA, et al. Feasibility of ambulatory, continuous 24-hour finger arterial pressure recording. Hypertension. 1993;21:65. [PubMed] [Google Scholar]

28. Ingelsson E, Bjorklund-Bodegard K, Lind L, et al. Diurnal blood pressure pattern and risk of congestive heart failure. Jama. 2006;295:2859. [PubMed] [Google Scholar]

29. Jhalani J, Goyal T, Clemow L, et al. Anxiety and outcome expectations predict the white-coat effect. Blood Press Monit. 2005;10:317. [PubMed] [Google Scholar]

30. Khattar RS, Senior R, Lahiri A. Cardiovascular outcome in white-coat versus sustained mild hypertension: a 10-year follow-up study. Circulation. 1998;98:1892. [PubMed] [Google Scholar]

31. Kikuya M, Hozawa A, Ohokubo T, et al. Prognostic significance of blood pressure and heart rate variabilities: the Ohasama study. Hypertension. 2000;36:901. [PubMed] [Google Scholar]

32. King GE. Errors in clinical measurement of blood pressure in obesity. Clin Sci. 1967;32:223. [PubMed] [Google Scholar]

33. King GE. Influence of rate of cuff inflation and deflation on observed blood pressure by sphygmomanometry. Am Heart J. 1963;65:303. [PubMed] [Google Scholar]

34. Kleinert HD, Harshfield GA, Pickering TG, et al. What is the value of home blood pressure measurement in patients with mild hypertension? Hypertension. 1984;6:574. [PubMed] [Google Scholar]

35. Lenfant C. Working group report on high blood pressure in pregnancy. J Clin Hypertens (Greenwich) 2001;3:75. [PMC free article] [PubMed] [Google Scholar]

36. Ma Y, Temprosa M, Fowler S, et al. Evaluating the accuracy of an aneroid sphygmomanometer in a clinical trial setting. Am J Hypertens. 2009;22:263. [PMC free article] [PubMed] [Google Scholar]

37. Mancia G, Bertinieri G, Grassi G, et al. Effects of blood-pressure measurement by the doctor on patient’s blood pressure and heart rate. Lancet. 1983;2:695. [PubMed] [Google Scholar]

38. Marey EJ. Physiologie Experimentale. Paris: 1876. Pression et vitesse du sang. [Google Scholar]

39. Mauck GW, Smith CR, Geddes LA, et al. The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure--part ii. J Biomech Eng. 1980;102:28. [PubMed] [Google Scholar]

40. Maxwell MH, Waks AU, Schroth PC, et al. Error in blood-pressure measurement due to incorrect cuff size in obese patients. Lancet. 1982;2:33. [PubMed] [Google Scholar]

41. McCubbin JA, Wilson JF, Bruehl S, et al. Gender effects on blood pressures obtained during an on-campus screening. Psychosom Med. 1991;53:90. [PubMed] [Google Scholar]

42. Mejia AD, Egan BM, Schork NJ, et al. Artefacts in measurement of blood pressure and lack of target organ involvement in the assessment of patients with treatment-resistant hypertension. Ann Intern Med. 1990;112:270. [PubMed] [Google Scholar]

43. Messelbeck J, Sutherland L. Applying environmental product design to biomedical products research. Environ Health Perspect. 2000;108 (Suppl 6):997. [PMC free article] [PubMed] [Google Scholar]

44. Mion D, Pierin AM. How accurate are sphygmomanometers? J Hum Hypertens. 1998;12:245. [PubMed] [Google Scholar]

45. Mitchell PL, Parlin RW, Blackburn H. Effect of Vertical Displacement of the Arm on Indirect Blood-Pressure Measurement. N Engl J Med. 1964;271:72. [PubMed] [Google Scholar]

46. Muller JE, Kaufmann PG, Luepker RV, et al. Mechanisms Precipitating Acute Cardiac Events: Review and Recommendations of an NHLBI Workshop. Circulation. 1997;96:3233. [PubMed] [Google Scholar]

47. Netea RT, Elving LD, Lutterman JA, et al. Body position and blood pressure measurement in patients with diabetes mellitus. J Intern Med. 2002;251:393. [PubMed] [Google Scholar]

48. Netea RT, Lenders JW, Smits P, et al. Both body and arm position significantly influence blood pressure measurement. J Hum Hypertens. 2003;17:459. [PubMed] [Google Scholar]

49. Netea RT, Smits P, Lenders JW, et al. Does it matter whether blood pressure measurements are taken with subjects sitting or supine? J Hypertens. 1998;16:263. [PubMed] [Google Scholar]

50. Nielsen PE, Janniche H. The accuracy of auscultatory measurement of arm blood pressure in very obese subjects. Acta Med Scand. 1974;195:403. [PubMed] [Google Scholar]

51. Niiranen TJ, Hanninen MR, Johansson J, et al. Home-measured blood pressure is a stronger predictor of cardiovascular risk than office blood pressure: the Finn-Home study. Hypertension. 2010;55:1346. [PubMed] [Google Scholar]

52. O’Brien E, Petrie J, Littler W, et al. The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens. 1990;8:607. [PubMed] [Google Scholar]

53. O’Brien E, Waeber B, Parati G, et al. Blood pressure measuring devices: recommendations of the European Society of Hypertension. BMJ. 2001;322:531. [PMC free article] [PubMed] [Google Scholar]

54. O’Callaghan WG, Fitzgerald DJ, O’Malley K, et al. Accuracy of indirect blood pressure measurement in the elderly. Br Med J (Clin Res Ed) 1983;286:1545. [PMC free article] [PubMed] [Google Scholar]

55. Ogedegbe G, Pickering TG, Clemow L, et al. The misdiagnosis of hypertension: the role of patient anxiety. Arch Intern Med. 2008;168:2459. [PMC free article] [PubMed] [Google Scholar]

56. Ohkubo T, Hozawa A, Yamaguchi J, et al. Prognostic significance of the nocturnal decline in blood pressure in individuals with and without high 24-h blood pressure: the Ohasama study. J Hypertens. 2002;20:2183. [PubMed] [Google Scholar]

57. Ohkubo T, Imai Y, Tsuji I, et al. Home blood pressure measurement has a stronger predictive power for mortality than does screening blood pressure measurement: a population-based observation in Ohasama, Japan. J Hypertens. 1998;16:971. [PubMed] [Google Scholar]

58. Ohkubo T, Imai Y, Tsuji I, et al. Prediction of mortality by ambulatory blood pressure monitoring versus screening blood pressure measurements: a pilot study in Ohasama. Journal of Hypertension. 1997;15:357. [PubMed] [Google Scholar]

59. Ohkubo T, Imai Y, Tsuji I, et al. Relation between nocturnal decline in blood pressure and mortality. The Ohasama Study. Am J Hypertens. 1997;10:1201. [PubMed] [Google Scholar]

60. Padfield PL, Jyothinagaram SG, Watson DM, et al. Problems in the measurement of blood pressure. J Hum Hypertens. 1990;4 (Suppl 2):3. [PubMed] [Google Scholar]

61. Parati G, Casadei R, Groppelli A, et al. Comparison of finger and intra-arterial blood pressure monitoring at rest and during laboratory testing. Hypertension. 1989;13:647. [PubMed] [Google Scholar]

62. Parati G, Pomidossi G, Casadei R, et al. Lack of alerting reactions to intermittent cuff inflations during noninvasive blood pressure monitoring. Hypertension. 1985;7:597. [PubMed] [Google Scholar]

63. Penaz J. Photo-electric measurement of blood pressure, volume and flow in the finger. Dresden: Digest of Tenth International Conference on Medical Biological Engineering; 1973. [Google Scholar]

64. Perloff D, Grim C, Flack J, et al. Human blood pressure determination by sphygmomanometry. Circulation. 1993;88:2460. [PubMed] [Google Scholar]

65. Perloff D, Sokolow M. Ambulatory blood pressure: mortality and morbidity. J Hypertens Suppl. 1991;9:S31. [PubMed] [Google Scholar]

66. Perloff D, Sokolow M, Cowan RM, et al. Prognostic value of ambulatory blood pressure measurements: further analyses. J Hypertens Suppl. 1989;7:S3. [PubMed] [Google Scholar]

67. Petrie JC, O’Brien ET, Littler WA, et al. Recommendations on blood pressure measurement. Br Med J (Clin Res Ed) 1986;293:611. [PMC free article] [PubMed] [Google Scholar]

68. Phillips RA, Diamond JA. Ambulatory blood pressure monitoring and echocardiography--Noninvasive techniques for evaluation of the hypertensive patient. Progress in Cardiovascular Diseases. 1991;41:397. [PubMed] [Google Scholar]

69. Pickering GW. High blood pressure. London: Churchill; 1968. [Google Scholar]

70. Pickering TG. Blood pressure variability and ambulatory monitoring. Curr Opin Nephrol Hypertens. 1993;2:380. [PubMed] [Google Scholar]

71. Pickering TG. The ninth Sir George Pickering memorial lecture Ambulatory monitoring and the definition of hypertension. Journal of Hypertension. 1992;10:401. [PubMed] [Google Scholar]

72. Pickering TG, Hall JE, Appel LJ, et al. Recommendations for blood pressure measurement in humans and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension. 2005;45:142. [PubMed] [Google Scholar]

73. Pickering TG, James GD, Boddie C, et al. How common is white coat hypertension? JAMA. 1988;259:225. [PubMed] [Google Scholar]

74. Pickering TG, Miller NH, Ogedegbe G, et al. Call to action on use and reimbursement for home blood pressure monitoring: executive summary: a joint scientific statement from the American Heart Association, American Society Of Hypertension, and Preventive Cardiovascular Nurses Association. Hypertension. 2008;52:1. [PubMed] [Google Scholar]

75. Pringle E, Phillips C, Thijs L, et al. Systolic blood pressure variability as a risk factor for stroke and cardiovascular mortality in the elderly hypertensive population. J Hypertens. 2003;21:2251. [PubMed] [Google Scholar]

76. Reder RF, Dimich I, Cohen ML, et al. Evaluating indirect blood pressure measurement techniques: a comparison of three systems in infants and children. Pediatrics. 1978;62:326. [PubMed] [Google Scholar]

77. Redon J, Campos C, Narciso ML, et al. Prognostic value of ambulatory blood pressure monitoring in refractory hypertension: a prospective study. Hypertension. 1998;31:712. [PubMed] [Google Scholar]

78. Sander D, Kukla C, Klingelhofer J, et al. Relationship between circadian blood pressure patterns and progression of early carotid atherosclerosis: A 3-year follow-up study. Circulation. 2000;102:1536. [PubMed] [Google Scholar]

79. Santucci S, Cates EM, James GD, et al. A comparison of two ambulatory blood pressure monitors, the Del Mar Avionics Pressurometer IV and the Spacelabs 90202. Am J Hypertens. 1989;2:797. [PubMed] [Google Scholar]

80. Sega R, Facchetti R, Bombelli M, et al. Prognostic value of ambulatory and home blood pressures compared with office blood pressure in the general population: follow-up results from the Pressioni Arteriose Monitorate e Loro Associazioni (PAMELA) study. Circulation. 2005;111:1777. [PubMed] [Google Scholar]

81. Shennan A, Gupta M, Halligan A, et al. Lack of reproducibility in pregnancy of Korotkoff phase IV as measured by mercury sphygmomanometry. Lancet. 1996;347:139. [PubMed] [Google Scholar]

82. Spence JD, Sibbald WJ, Cape RD. Direct, indirect and mean blood pressures in hypertensive patients: the problem of cuff artefact due to arterial wall stiffness, and a partial solution. Clin Invest Med. 1979;2:165. [PubMed] [Google Scholar]

83. Staessen JA, Thijs L, Fagard R, et al. Predicting cardiovascular risk using conventional vs ambulatory blood pressure in older patients with systolic hypertension. Systolic Hypertension in Europe Trial Investigators. Jama. 1999;282:539. [PubMed] [Google Scholar]

84. van Egmond J, Hasenbos M, Crul JF. Invasive v. non-invasive measurement of arterial pressure. Comparison of two automatic methods and simultaneously measured direct intra-arterial pressure. Br J Anaesth. 1985;57:434. [PubMed] [Google Scholar]

85. van Egmond J, Lenders JW, Weernink E, et al. Accuracy and reproducibility of 30 devices for self-measurement of arterial blood pressure. Am J Hypertens. 1993;6:873. [PubMed] [Google Scholar]

86. van Montfrans GA, van der Hoeven GM, Karemaker JM, et al. Accuracy of auscultatory blood pressure measurement with a long cuff. Br Med J (Clin Res Ed) 1987;295:354. [PMC free article] [PubMed] [Google Scholar]

87. Veerman DP, van Montfrans GA, Karemaker JM, et al. Inflating one’s own cuff does not increase self-recorded blood pressure. J Hypertens Suppl. 1988;6:S77. [PubMed] [Google Scholar]

88. Veerman DP, van Montfrans GA, Wieling W. Effects of cuff inflation on self-recorded blood pressure. Lancet. 1990;335:451. [PubMed] [Google Scholar]

89. Verdecchia P, Angeli F, Gattobigio R, et al. Impact of blood pressure variability on cardiac and cerebrovascular complications in hypertension. Am J Hypertens. 2007;20:154. [PubMed] [Google Scholar]

90. Verdecchia P, Schillaci G, Borgioni C, et al. Prognostic value of left ventricular mass and geometry in systemic hypertension with left ventricular hypertrophy. The American Journal of Cardiology. 1996;78:197. [PubMed] [Google Scholar]

91. Verdecchia P, Schillaci G, Borgioni C, et al. White coat hypertension and white coat effect. Similarities and differences. Am J Hypertens. 1995;8:790. [PubMed] [Google Scholar]

92. Villar J, Repke J, Markush L, et al. The measuring of blood pressure during pregnancy. Am J Obstet Gynecol. 1989;161:1019. [PubMed] [Google Scholar]

93. Villegas I, Arias IC, Botero A, et al. Evaluation of the technique used by health-care workers for taking blood pressure. Hypertension. 1995;26:1204. [PubMed] [Google Scholar]

94. Webster J, Newnham D, Petrie JC, et al. Influence of arm position on measurement of blood pressure. Br Med J (Clin Res Ed) 1984;288:1574. [PMC free article] [PubMed] [Google Scholar]

95. Wonka F, Thummler M, Schoppe A. Clinical test of a blood pressure measurement device with a wrist cuff. Blood Press Monit. 1996;1:361. [PubMed] [Google Scholar]

96. Yarows SA, Qian K. Accuracy of aneroid sphygmomanometers in clinical usage: University of Michigan experience. Blood Press Monit. 2001;6:101. [PubMed] [Google Scholar]

97. Yelderman M, Ream AK. Indirect measurement of mean blood pressure in the anesthetized patient. Anesthesiology. 1979;50:253. [PubMed] [Google Scholar]


Page 2

Cuff sizes recommended by the American Heart Association

CuffArm circumference (cm)Bladder width (cm)Bladder length (cm)
Newborn<636
Infant6–15515
Child16–21821
Small adult22–261024
Adult27–341330
Large adult35–441638
Adult thigh45–522042