Specifications Model Milking Control Units Metatron, DeMax & DemaTron Documents 100% found this document useful (1 vote) 2K views 4 pages Description:Daftar Error YP90-100 Copyright© © All Rights Reserved Available FormatsDOC, PDF, TXT or read online from Scribd Share this documentDid you find this document useful?100% found this document useful (1 vote) 2K views4 pages GEA YP90-YP100 Error List & Trobleshooting
TABLE 3.1 ALARM (CONTINUE)ALARM MESSAGE DESCRIPTIONCode H0.1 ROM inside of main MCU fails. At this time, the heater will stop working ,press Silence/Reset Key to clear alarming for 4min, please refer to theservice part.Code H0.2 Internal system inside of main MCU fails. At this time, the heater will stopworking, press Silence/Reset Key to clear alarming for 4min, please refer to the service part.Code H0.3 Communication of main and auxiliary MCU A fails, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min,please refer to the service part.Code H0.4 ROM inside of auxiliary MCU fails, At this time, the heater will stopworking, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.Code H0.5 Internal system inside of auxiliary MCU fails, At this time, the heater willstop working, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.Code H0.6 E2ROM inside main MCU fails, At this time, the heater will stop working ,press Silence/Reset Key to clear alarming for 4min, please refer to theservice part.Code H0.7 SRAM failure (including system cord failure). At this time, the heater willstop working, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.Code H0.8 Main ADC failure(e.g., auxiliary system TLV2548, including wrong sample,chip damage, no response and so on). At this time, the heater will stopworking, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.Code H0.9 Auxiliary ADC failure(e.g., auxiliary system TLV2544, including wrongsample, chip damage, no response and so on). At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min,please refer to the service part.Code H1.0 Heat circuit failure(e.g., heater, solid relay, mechanical relay and allcontrol circuit fails). At this time, the heater will stop working, pressSilence/Reset Key to clear alarming for 4min, please refer to the servicepart.Code H1.1 The timer RTC inside the system fails or is not accurate, At this time, theheater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part.Code H1.2 If the keyboard circuit is broken, At this time, the heater will stop working,press Silence/Reset Key to clear alarming for 4min, please refer to theservice part.The system indicator is onCode H1.3 If the battery inside of the controller is error, At this time, the heater willstop working, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.PROCEDURE
At first, connected power supply, press and hold Set key for 2s as soon as switch on untilAirtemprature indicator displays “ --.-“, and no display on Skin and Set temprature indicator. And then,press Up or Down until Air temprature indicator displays “78.6”, after then, pressSet key to enter intothe system set screen, or press Silence/Reset key to exit.After entering into the set screen, Set temprature indicator displays set code, Air temprature indicator displays the specific set data.Turn on the power switch5-3 I2Press and hold the for 3sEnter passwordPress key, EXIT Press or key to enter password 78.6,Then press key to confirm Error Password checkYES System setting screenPress key or Key, set temperature indicator displaysthe set codePress or Key can change current set value ENDPress key Press and together SAVE and EXIT Return to original set Brightness set of indicator SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0 CODE 00.1LED Lightness setIndicate the lightness , and set range is 00.1~01.5. The larger the lighter.CODE 00.2 Air temperature sensor compensation Set value of temperature compensation for air temperature sensor , and set range is -5.0~05.0 ,the first figure of set value means the increasing and decreasing value. For Example, 00.2Means that the air temperature adds 0.2 as compensation. -1.3 means that the air temperature reduces as compensation. CODE 00.3 Air flow temperature sensor compensationtemperature compensation for air flow temperature sensor , and set range is -5.0~05.0 , thefirst figure of set value means the increasing and decreasing value. For Example, 00.2means thatthe air flow temperature adds 0.2 ascompensation. -1.3means that the air flow temperaturereduces 1.3 ascompensation.CODE 00.4 Skin temperature sensor compensationSet value of temperature compensation for skin temperature sensor , and set range is-5.0~05.0 , the first figure of set value means the increasing and decreasing value. For Example,00.2means that the skin temperature adds 0.2 as compensation. -1.3means that the skintemperature reduces 1.3 ascompensation.CODE 00.5 Machine Number Set method of machine number, and set range is 00.0~02.0. For example, 00.1 measn No. 1machine, 01.5 means No. 15.CODE 00.6 Motor speed alarm valueSet method of fan speed, and set range is 00.0~25.5. Every 0.1 means 20 rpm. For example,05.0 means 1000 rpm. Alarm will be active if the fan speed is lower than this value.CODE 00.7 Set value of Low deviation for air temperatureSet method of low deviation alarm in air mode, and set range is 00.0~05.0. For example, 00.8means 0.8 . CODE 00.8 Set value of High deviation for air temperatureSet method of high deviation alarm in air mode, and set range is 00.0~05.0. For example, 00.8means 0.8 .CODE 00.9 Set value of low deviation for skin temperature
Set method of low deviation alarm in skin mode, and set range is 00.0~05.0. For example, 00.8means 0.8 . CODE 01.0 Set value of high deviation for skin temperatureSet method of high deviation alarm in skin mode, and set range is 00.0~05.0. For example, 00.8means 0.8 .CODE 01.1 Minimum set value of air temperatureSet method of Minimum set value of air temperature, and set range is 20.0~34.0. For example,25.0 means Minimum set value 25.0 . CODE 01. 2 Maximum set value of air temperatureSet method of Maximum set value of air temperature in mode of>37.0, and set range is37.0~39.0. For example, 39.0 means Maximum set value 39.0 . CODE 01.3 Minimum set value of skin temperatureSet method of Minimum set value of skin temperature, and set range is 30.0~34.0. For example,34.0 means Minimum set value 25.0 . CODE 01.4 Maximum set value of skin temperatureSet method of Maximum set value of skin temperature in mode of>37.0, and set range is37.0~39.0. For example, 39.0 means Maximum set value 39.0 . CODE 01.5 Set value of Over temperature alarm A 37 Set method of Over temperature alarm in air mode (set temperatuer 37, and set range is37.5~39.0. For example, 38.0 means over-temp alarm value 38.0 . CODE 01.6 Set value of Over temperature alarm B ( 37 )Set method of Over temperature alarm in air mode (set temperatuer 37, and set range is38.5~40.0. For example, 40.0 means over-temp alarm value 40.0 . CODE 01.7 Set value of air flow Over temperature alarm A (37Set method of alarm value for air flow blocked (set temperatuer 37, and set range is38.5~48.0. For example, 45.0 means alarm value for air flow blocked 45.0 CODE 01.8Set value of air flow Over temperature alarm B (37)Set method of alarm value for air flow blocked (set temperatuer >37, and set range is 40.5~52.0.For example, 45.0 means alarm value for air flow blocked 45.0 5.4.2 TEST EQUIPMENT REQUIREDThe test equipment listed below is required for troubleshooting the Controller. Equivalent testequipment may be substituted.Digital VOM, FLUKE 8060A31.0Air Flow Temperature SimulationResistor R=23166 Ω / 0.25W/0.5%TABLE 5.1 ERROR CODES |